Tilt angle specification
|
Tilt to
|
Always positive!
|
End at
|
Always negative!
|
Run Series in two directions
|
Always selected, set pre-tilt for lamellae.
|
Delay time after tilting
|
3 seconds for now, might change in th future.
|
Use dose-symmetric scheme
|
Always selected.
|
Dose-symmetric Setup
|
Scroll down this page.
|
Size, Image Shift and recentering Specifications
|
Limit Image shift to
|
15 um at start, dont change!
|
Nothing else selected.
|
|
Beam Intensity or Exposure Control
|
Keep beam intensity constant
|
Always on.
|
| Nothing else selected. |
|
Camera and Acquisition Parameters
|
EF-CCD
|
Selected.
|
| Nothing else selected. |
|
Autofocus Control
|
Defocus target
|
Set for each position untill we have an automated solution from David.
|
Beam Tilt
|
4 mrad, 10 mrad for VPP Tomo.
|
Iteration threshold
|
0.1 um for high-resolution subtomo, something else for other Tomo: how much focus spread through a tiltseries is ok?
|
Do autofocus at least every x degrees
|
0.1
um for high-resolution subtomo (autofocus each tilt), something else
for other Tomo: how much focus spread through a tiltseries is ok? |
Stop if autofocus change exceeds
|
10 um seems legit....
|
Initial and Final Actions
|
Nothing selected!
|
|
Tracking Control Parameters
|
Wait for drift to settle
|
Not needed anymore on K3 and we use a 3 sec wait after tilt time allready.
|
Get tracking image when error in X/Y prediction is > X % of image size
|
0 % for high-resolution subtomo
(track each tilt), something else (the default 5%?) for other Tomo: how
much image shift between tilts through a tiltseries is ok? |
Track
|
Before
|
Align only with tracking images, not Record images
|
Do never select this!
|
Align with Preview befor getting new track reference
|
??????? don't know.
|
Get new track reference if record alignment differs by > X %
|
??????? don't know. |